TY - JOUR AB - Background and purpose — Temporary hemiepiphysiodesis for growth modulation in skeletally immature patients is a long-known technique. Recently the use of tension-band devices has become popular. This study compares 2 tensionband implants (eight-Plate and FlexTack) regarding their effects on the growth plate. Animals and methods — 12 pigs in 2 equally sized groups (A and B) were investigated. The right proximal medial tibia was treated with either eight-Plate or FlexTack. The left tibia of the same pig was treated with the opposite implant. After 9 weeks all implants were removed. Animals in group B were then hosted for another 5 weeks. Histomorphometric analysis of the growth plate was carried out after 9 and 14 weeks, respectively. Radiographs were taken at implantation, removal, and after 14 weeks. Results — Both tension-band devices achieved a statis- tically significant and clinically relevant growth inhibition, whereas the effect appeared to be more distinct after the use of FlexTack. Implant-related complications or physeal damage was not observed. After implant removal, rebound phenomenon was radiologically observed in all cases. The growth plates treated with eight-Plate showed a paradox reversal of the zonal distributions, with an increase of the proliferative zones at the previously arrested medial aspect of the physis and a decrease laterally. Interpretation — Both eight-Plate and FlexTack proved to be appropriate devices for growth-guiding treatment. The radiographic evaluation showed a change in angular axes after treatment with each implant, while the correction appeared to be faster with FlexTack. The paradox cartilaginous reaction observed after removal of the eight-Plate might be a histopathological correlate for rebound phenomenon. AU - Laufer, Andrea Maria AU - Sattelberger, Julia AU - Hillebrand, Hauke AU - Gosheger, Georg AU - Frommer, Adrien AU - Frommer, Adrien Tobias AU - Appelbaum, Sebastian AU - Abood, Ahmed Abdul-Hussein AU - Gottliebsen, Martin AU - Rahbek, Ole AU - Møller-Madsen, Bjarne AU - Rödl, Robert AU - Vogt, Björn DA - 2021-01-19 DO - 10.17879/42099686425 LA - eng N1 - Acta Orthopaedica 92 (2021) 3, 364-370 N1 - Finanziert durch den Open-Access-Publikationsfonds der Westfälischen Wilhelms-Universität Münster (WWU Münster). PY - 2021-01-19 TI - Comparison of histomorphometric and radiographic effects of growth guidance with tension-band devices (eight-Plate and FlexTack) in a pig model UR - https://nbn-resolving.org/urn:nbn:de:hbz:6-22099656460 Y2 - 2024-11-22T04:18:45 ER -